Siemens Healthineers: OPTIQ AI on ARTIS angiography systems, exposure control based on CNR and noise reduction with artificial intelligence

The OPTIQ AI imaging chain regulates exposure based on the contrast-to-noise ratio (CNR) to maintain image quality according to the ALARA principle. It includes AI algorithms for noise reduction, exposure optimisation for metallic materials, and advanced functions such as DSA Roadmap.

27 November 2026 · Source: siemens-healthineers.com · 2 min read

This English text is an automatic translation of the Italian original, which remains the reference. Read the original in Italian.

In angiographic systems, image quality tends to deteriorate with steep C-arm angles and in large patients due to increased beam path thickness. Conventional Automatic Exposure Control (AEC) adjusts parameters to maintain a constant dose at the detector, not image quality. The OPTIQ imaging chain by Siemens Healthineers, in the OPTIQ AI version described on the manufacturer's Italian page, instead uses the contrast-to-noise ratio (Contrast-to-Noise Ratio - CNR) as the control parameter. Manufacturer-reported experiences concern angiographic systems from the ARTIS family (ARTIS icono and ARTIS genio).

  • Exposure adjustment "5+1": tube voltage, current, filtration, focal spot size, pulse width, and detector dose are adjusted in real time, even during system movement, to maintain a preset CNR at the lowest reasonably achievable dose (ALARA principle - As Low As Reasonably Achievable). The manufacturer specifies that quality constancy holds within the system's physical limits.
  • Noise reduction with artificial intelligence (Artificial Intelligence - AI): an algorithm acts in real time on various 2D imaging modes and anatomical regions. The improvement is declared compared to the noise reduction function of the previous software version (VE40) and was evaluated via non-clinical tests on phantom images; structure preservation, such as devices, was measured using the Structural Similarity Index Measure (SSIM).
  • Structure Scout: exposure is optimised for device material (e.g., iron, tantalum, or platinum) based on its X-ray absorption. In the manufacturer's clinical examples regarding individual patients, the Dose Area Product (DAP) per frame drops from 0.63 to 0.45 µGy·m² in a TIPS procedure (transjugular intrahepatic portosystemic shunt) and from 0.15 to 0.026 µGy·m² in a treatment of arteriovenous malformation with tantalum embolic agent, with equivalent device visibility.
  • Overlay Ref: increases vascular contrast in the reference image in Digital Subtraction Angiography (DSA) overlaid on fluoroscopy, while maintaining device visibility.
  • DSA Roadmap: allows reusing any previously acquired DSA scene, with automatic system return to the same position and a declared precision of less than 0.5 mm, with potential dose and contrast medium reduction.
  • DSA acquisition: three-focal spot tube, 2K live images, up to five magnification formats, and real-time pixel shifting in six dimensions.

Regulatory status: OPTIQ AI is already commercially available in Europe as a medical device on new ARTIS angiographic systems; the manufacturer reports clinical experiences from Germany (Kerckhoff-Klinik in Bad Nauheim). In the United States, six new interventional systems with the OPTIQ AI chain received Food and Drug Administration (FDA) 510(k) authorisation in May 2026. The note on the manufacturer's Italian page, which states the product is not yet on the market, is now outdated. Dose reduction data derive from phantom measurements and individual cases provided by the manufacturer. For the previous OPTIQ chain (CNR-based control and material-specific imaging), the manufacturer cites scientific publications on ARTIS systems.