IQ Endoscopes: indication extension to flexible sigmoidoscopy for portable single-use endoscopic system
The British company IQ Endoscopes has received regulatory approval to expand the indications for its single-use endoscopic system to include flexible sigmoidoscopy, in addition to the already authorised gastroscopy. The platform enables gastrointestinal procedures to be performed outside a dedicated endoscopy suite, eliminating the need for reprocessing.
This English text is an automatic translation of the Italian original, which remains the reference. Read the original in Italian.
On 20 November 2026, IQ Endoscopes, a British company based in Chepstow (Wales), announced a regulatory approval extending indications for its single-use digestive endoscopy platform to flexible sigmoidoscopy, whereas previously it was indicated only for gastroscopy. This news appears in an article by MedTech World declared by the publication as sponsored content; therefore, it should be read as corporate communication, not independent journalism. The article does not specify which authority issued the approval (FDA, UKCA or CE marking) nor for which market; it generically cites the growing demand for gastrointestinal diagnostics in Europe.
Statements attributed to the company:
- A system for multiple procedures: a single portable system for multiple gastrointestinal applications, simpler to distribute across different care settings.
- Endoscopy outside the suite: suitable cases are performed outside the endoscopy suite, freeing up rooms for more complex interventions.
- No reprocessing: with single-use devices, reprocessing, maintenance and repairs disappear, and the device is immediately available.
- Adjunct, not replacement: the system is designed to complement reusable endoscopes, not to replace them.
The already authorised system: according to the FDA summary of 510(k) K232028 (17/11/2023, Class II, product code FDS, 21 CFR 876.1500), the Q Vision G-100 gastroscope is flexible, sterile and single-use, with a distal CMOS sensor, integrated LEDs, an operating channel, and channels for air, water and suction; it is indicated for examination of the oesophagus, stomach and first and second portions of the duodenum in adults. It operates only with the reusable base unit Q Base Q-100, which provides power, compressed air and data via a proprietary connector and sends video to an external medical monitor via HDMI or BNC. The base unit has no software, user menus or image processing functions such as BLI or LCI of the predicate device. The G-100 received UKCA approval in November 2023; the company raised £4 million in 2021 (Series A) and £5.2 million in 2023 (Series B).
Single-use between infection safety and environmental footprint: the strongest evidence regarding risks linked to reprocessing concerns duodenoscopes, not sigmoidoscopy instruments. Since 2019, the FDA has recommended switching to duodenoscopes with single-use components or entirely single-use devices, because many steps in reprocessing instructions cannot be reliably performed. On the environmental side, the ESGE-ESGENA position statement of 2022 advises against routine use of single-use endoscopes, reserving them for highly selected patients, and estimates that widespread use would increase net waste mass by approximately 40%; a 2024 review (J Gastroenterol Hepatol) confirms this stance. Therefore, authorisation of a device is not sufficient to justify its extended use: it is necessary to define in which scenarios single-use provides a real benefit.
Aspects for HTA evaluation and clinical engineering:
- Cost per procedure: comparison between single-use and the full cost of reusable devices (depreciation, reprocessing with endoscope washers, chemicals, personnel and traceability, maintenance, repairs, downtime).
- Use scenarios: procedures at the patient's bedside, in emergency departments or in sites without a central reprocessing unit; continuity of service during downtimes. No published clinical data on sigmoidoscopy with this system are available.
- Image performance: the base unit has no image processing functions; it must be evaluated whether the image quality is adequate for the clinical indication.
- Integration: proprietary connector and video output HDMI/BNC to an external monitor; compatibility with existing image acquisition and reporting systems must be verified.
- Waste and sustainability: estimation of additional healthcare waste volumes, to be compared with the ESGE-ESGENA position.
- Regulatory status in Italy: before any purchase, CE marking according to Regulation (EU) 2017/745 must be verified; the sources found cite only UKCA and FDA.
What is still unknown: which authority issued the November 2026 approval, on what date and for which market; whether sigmoidoscopy is performed with a dedicated new endoscope or with the G-100 under extended indication; whether published clinical studies exist on performance in sigmoidoscopy; whether the product has CE marking and is distributed in Italy. Given that this is sponsored content, it is advisable to seek confirmation from independent sources (FDA database, UKCA/MHRA registers, EUDAMED) or an official company press release.
Text by the HTS editorial team.
Primary source: https://med-tech.world/news/iq-endoscopes-reimagining-endoscopy-with-single-use-innovation/
Other sources: accessdata.fda.gov; accessdata.fda.gov; iqendoscopes.com; iqendoscopes.com; iqendoscopes.com; fda.gov; healio.com; pubmed.ncbi.nlm.nih.gov; onlinelibrary.wiley.com.
Related news
EndoscopyFujifilm presents the ELUXEO system with OXEI technology for dye-free tissue perfusion assessment
The new ELUXEO surgical system, equipped with OXEI hyperspectral imaging technology, enables surgeons to monitor oxygen saturation in superficial tissues during endoscopic procedures without the use of fluorescent dyes or contrast agents.